Chromperfect Instrument Setpoint Files: How to Upload, Open and Edit Them Correctly

Writer: Chromperfect
Chromperfect
Aug 1
5 min read

Watch the video to learn how to upload, open and edit Chromperfect instrument setpoint files correctly, including MIC and instrument-specific editors, the Awaiting Download status, and the essential rule: upload first, then edit.

Instrument setpoint files are essential when Chromperfect is used to control a digitally connected chromatograph. They contain the operating conditions that are downloaded to the instrument, including settings such as temperatures, pressures, flows, detector parameters and other instrument-specific controls.


Although the basic workflow is straightforward, new users often encounter confusion because instrument setpoint files behave differently from many other Chromperfect file types. Users may search for a separate setpoint editor, attempt to create a blank setpoint file, or try to upload settings while the instrument is still showing Awaiting Download.


This guide explains the correct workflow and the terminology used within Chromperfect.


What are instrument setpoint files?


Instrument setpoint files contain the settings used by Chromperfect to control a supported digital instrument.


The exact settings available depend on the chromatograph and the type of digital connection being used. A gas chromatograph may include controls for:

  • oven temperature programs

  • inlet temperatures and pressures

  • carrier-gas flow

  • detector temperatures

  • detector gases

  • valve events

  • signal selection

  • run time and sampling settings, where applicable


Because different instrument families expose different controls, Chromperfect does not use one universal editor layout for every instrument. Each supported instrument family or control interface has its own setpoint file format and its own editing form.


Understanding the Instrument Setpoint File Editor


Users may hear references to the Method File Editor, Calibration File Editor, Format File Editor or Setpoint File Editor.


The last term can be confusing because a setpoint file is not simply another processing file. It is the file used to control the physical instrument. For clarity, the more precise term is:


Instrument Setpoint File Editor


Within the Instrument Setpoint File Editor, Chromperfect provides different forms for different supported instruments and control interfaces. Each form reflects the controls available on that particular instrument.


For example, users may hear references to:

  • the Agilent 7890 Setpoint File Editor

  • the MIC Editor for ICF or RC.net instruments

  • the PerkinElmer setpoint editor

  • the Thermo instrument setpoint editor

  • other instrument-specific editor names


These terms do not normally refer to separate programs installed on the Windows desktop. They describe the instrument-specific form that opens inside the Chromperfect File Editor.



Chromperfect File Editor showing the File menu used to open an instrument setpoint file.
Instrument setpoint files are opened through the main Chromperfect File Editor.

One File Editor, multiple editing forms


The Chromperfect File Editor is the central program used to open and edit Chromperfect file types.


When a file is opened, Chromperfect displays the appropriate editing form:

  • opening a Method file displays the Method Editor

  • opening a Calibration file displays the Calibration Editor

  • opening a Format file displays the Format Editor

  • opening an instrument setpoint file displays the appropriate instrument-specific setpoint editor


Therefore, when someone says, “Open the setpoint editor,” the practical instruction is usually:

  1. Open the Chromperfect File Editor.

  2. Select File > Open.

  3. Choose the appropriate setpoint file type.

  4. Open the required instrument setpoint file.


The name used in conversation may depend on the instrument. Someone working with an Agilent 7890 may refer to the 7890 Setpoint File Editor, while an ICF user may refer to the MIC Editor. Both are accessed through the same Chromperfect File Editor.


ICF and MIC instrument setpoint files


For instruments controlled through the Agilent Instrument Control Framework, also known as ICF or RC.net, the Chromperfect setpoint file normally uses the .MIC extension.

The MIC file is the ICF instrument setpoint file.


Internally, it contains the instrument settings used by the ICF control system. The ICF Setpoint Editor does not connect directly to the chromatograph. It opens and edits an existing MIC file through the Chromperfect File Editor.


To open an ICF setpoint file:

  1. Open Chromperfect File Editor.

  2. Choose File > Open.

  3. Select the setpoint file option.

  4. Locate and open the required MIC file.

  5. Wait for the ICF editor form to initialise.


ICF forms may take several seconds to initialise. During this time, the File Editor may temporarily appear unresponsive. This does not necessarily indicate a software problem.


Chromperfect File Editor opening a MIC instrument setpoint file for an ICF-controlled chromatograph.
MIC files are ICF instrument setpoint files and are opened through Chromperfect File Editor.

Why instrument setpoint files are not normally created as blank files


A common mistake is to treat an instrument setpoint file like a Method, Calibration, Format or Sequence file and attempt to create a new blank file inside File Editor.

Instrument setpoint files are different because they are tied to the physical instrument and its current hardware configuration.


The first setpoint file should normally be created by uploading the current settings from the connected chromatograph. This ensures that the file reflects the instrument that is actually connected, including its installed inlets, detectors, valves and other hardware options.


The correct rule is:


Upload first, then edit.


The normal workflow is:

  1. Connect Chromperfect to the instrument.

  2. Upload the current instrument settings.

  3. Save the uploaded setpoint file.

  4. Open that file in Chromperfect File Editor.

  5. Review or modify the settings.

  6. Save the file or use Save As to create another version.



Workflow showing a connected instrument uploading its setpoints, saving the file and opening it in Chromperfect File Editor.
The first instrument setpoint file should normally be uploaded from the connected chromatograph.

What to do when the instrument shows Awaiting Download


Another common difficulty occurs when the instrument status is Awaiting Download.

Awaiting Download means that Chromperfect has not yet successfully downloaded a Method to the instrument. The system is therefore not ready to begin a normal acquisition.


In this state, users may become stuck when attempting to upload instrument setpoints.

The practical solution is normally to download a simple Method that does not already reference a setpoint file. The purpose of this download is not to create the final analytical method. It is simply to initialise the instrument connection and move the status from Awaiting Download to Ready.


The workflow is:

  1. Open Chromperfect Acquisition.

  2. Check the instrument status.

  3. If the status is Awaiting Download, download a simple Method without an attached setpoint file.

  4. Confirm that the status changes to Ready.

  5. Upload the current setpoints from the instrument.

  6. Save the uploaded setpoint file.

  7. Open the file in Chromperfect File Editor.

  8. Edit and save the file as required.



Chromperfect Acquisition instrument status showing Ready after a Method download.
To apply an instrument control setpoint file the status should be READY

The correct instrument setpoint file workflow


The complete process can be summarised as follows:

Awaiting Download→ Download a simple Method without a setpoint file→ Confirm Ready→ Upload the current instrument setpoints→ Save the setpoint file→ Open it in Chromperfect File Editor→ Edit and save, or use Save As


This workflow avoids three common mistakes:

  • searching for a separate instrument setpoint editor program

  • trying to create the first setpoint file as a blank file

  • attempting to upload settings before the instrument reaches Ready


Key points to remember


The Chromperfect File Editor is the normal starting point for editing Chromperfect files.

The term Instrument Setpoint File Editor describes the group of instrument-specific setpoint editor forms available through File Editor.


Different instruments use different file formats, forms and layouts, so users may hear names such as MIC Editor, 7890 Setpoint File Editor or PerkinElmer Setpoint File Editor.

These are not normally separate desktop applications. They are all accessed through the Chromperfect File Editor.


The first instrument setpoint file should normally be uploaded from the connected instrument, not created as a blank file.


When an instrument shows Awaiting Download, first download a simple Method without a setpoint file and confirm that the instrument reaches Ready.


The practical rule remains:


Upload first, then edit.